RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done

HD-DIAL-252_Request Create Dialin Command TO Enable Air Bubble Alarm IN Service Mode Air Bubble
HD-DIAL-252_Request Create Dialin Command TO Enable Air Bubble Alarm IN Service Mode Air Bubble
Correct, no longer required. Removed.

Correct, no longer required. Removed.

Reset proper flag.

Reset proper flag.

We are (I think now) allowing door open in first 2 states (UF and invert dialyzer), so should not be getting door open alarm when UI is prompting user to invert dialyzer. When we get to 3rd state (...

We are (I think now) allowing door open in first 2 states (UF and invert dialyzer), so should not be getting door open alarm when UI is prompting user to invert dialyzer. When we get to 3rd state (where we start requiring door to be closed), we are looking for user to connect patient and confirm. UI should be following along with our states and showing appropriate screens.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Doesn't look removed. I'm talking about the call to handlePatientConnectionDoorOpened(). Do we need this now that door closed is required (as set in prior state)?

Doesn't look removed. I'm talking about the call to handlePatientConnectionDoorOpened(). Do we need this now that door closed is required (as set in prior state)?

Doesn't make sense to reset in previous state. You're resetting door opened flag there. I'm talking about patient connect confirm flag used in if statement above.

Doesn't make sense to reset in previous state. You're resetting door opened flag there. I'm talking about patient connect confirm flag used in if statement above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Updated to do this. To me it seems counter intuitive. If the user opens door when UF or BP screen is showing, then Open door screen come up, user opens door and gets alarm. If they hit next, it sho...

Updated to do this. To me it seems counter intuitive. If the user opens door when UF or BP screen is showing, then Open door screen come up, user opens door and gets alarm. If they hit next, it shows flip dialyzer screen but still alarms if door opened.

Removed

Removed

Removed

Removed

Reset flag in previous state.

Reset flag in previous state.

Removed

Removed

Done. Reset door flag also.

Done. Reset door flag also.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.