Merge branch 'DEN-15317-usability-hd-door-open-alarm-during-post-treatment' into develop

DEN-15317: Addressed DENF-15317.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

A 10 second timeout timer, driven by the 50ms tick that drives the alarm state machine.

A 10 second timeout timer, driven by the 50ms tick that drives the alarm state machine.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

HD-DEN-15451_Dpi Pump Does Not Resume After Door Opened Alarm Cleared IN Pre TX
HD-DEN-15451_Dpi Pump Does Not Resume After Door Opened Alarm Cleared IN Pre TX
FIXED IN CODE WALKTHROUGH

FIXED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Merge branch 'DEN-15451-dpi-pump-does-not-resume-after-door-opened-alarm-cleared-in-pre-tx' into develop

Add alarmsBlockedTimer to header. Is alarmsBlockedTimer a timer or is it a counter?

Add alarmsBlockedTimer to header. Is alarmsBlockedTimer a timer or is it a counter?

DIAL-258: code review changes

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Add current4thLevleState to header.

Add current4thLevleState to header.

Add current4thLevelState to header

Add current4thLevelState to header

DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power-4

Clear early initiation of block on cpld signal, HW recovery on release of cpld, SW recovery on alarm clear, normal alarm operation 10 seconds later.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15196-DIAL-258' into develop

DEN-15353: explicit float casting of treatment duration parameter when calculating max volume.

DEN-15451 Fix re-enable of DPi on alarm resume.

Merge branch 'staging' into DEN-15196-DIAL-258

HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power 4 - Ready
HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power 4 - Ready
HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power
HD-DEN-15306_HD Fault Valve Position Alarm After Restoring AC Power
compiler fixes