Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15196-DIAL-262-fixes' into develop

DIAL-262: removed old 4th state data when hd pauses in priming

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Add venLowPresExemptAfterAirTrapFillTimerCtr to header.

Add venLowPresExemptAfterAirTrapFillTimerCtr to header.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15558-fw-system-does-not-widen-venous-pressure-range-when-filling-air-trap-vbt-open-in' into develop

DEN-15558: Extend exemption for venous low pressure alarm during air trap fill for 5 seconds after fill to allow pressure to recover.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15363-fw-handle-bld-sensor-communication-errors' into staging

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Was temporary change for blood leak test. Restored original code.

Was temporary change for blood leak test. Restored original code.

DEN-15558: Finished restore of original BP speed during dialyzer prime.

DIAL-262: fixed operation status event to have more accurate transitioning in between dialysis sub modes.

DEN-15558: Reversed temporary change in dialyzer prime BP speed. Increased max venous pressure from 300 to 400 mmHg.

Replacing the calculation with a #define and the numeric value for BP flow rate might be helpful.

Replacing the calculation with a #define and the numeric value for BP flow rate might be helpful.

HD-DEN-15558_FW System Does Not Widen Venous Pressure Range When Filling Air Trap Vbt Open IN
HD-DEN-15558_FW System Does Not Widen Venous Pressure Range When Filling Air Trap Vbt Open IN
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflict.

DEN-15558: temporary slow down of BP during pre-tx dialyzer prime state to see if helps with blood leak issue.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo