I don't think you need to repeat @details Inputs: and @details Outputs:. Only need those for first row. See other functions for example.

I don't think you need to repeat @details Inputs: and @details Outputs:. Only need those for first row. See other functions for example.

Blank line before banner comment.

Blank line before banner comment.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Still more here than should be. Trimmer heater is being turned off twice now. DPo should already be stopped. And VBA/VBV should be handled by state machine I would think unless user stop requiremen...

Still more here than should be. Trimmer heater is being turned off twice now. DPo should already be stopped. And VBA/VBV should be handled by state machine I would think unless user stop requirement talks about valves too - and even if so, valves should only close when user stop alarm is active - nothing to do with recirc properties in alarms.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Change postponed for further testing. No change in BloodPumps. Done.

Change postponed for further testing. No change in BloodPumps.
Done.

Change postponed for further testing. No change in DialIn. Done.

Change postponed for further testing. No change in DialIn.
Done.

Change postponed for further testing. No change in DialOut. DONE.

Change postponed for further testing. No change in DialOut.
DONE.

Done.

Done.

Done.

Done.

BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. Removed anything that is off, and stays off in recirculation. Done.

BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. Removed anything that is off, and stays off in recirculation.
Done.

s/b recirculate. Done.

s/b recirculate. Done.

Done

Done

Done.

Done.

Agree that BP is on in the recirc state. Forget 2nd comment - focus on multiple function calls comment.

Agree that BP is on in the recirc state. Forget 2nd comment - focus on multiple function calls comment.

In treatment recirc, BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. It seems as though the user stop should turn off both pumps.

In treatment recirc, BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. It seems as though the user stop should turn off both pumps.

HD-DEN-15927_Alarm 127 Low Venous Pressure Triggers Post Treatment
HD-DEN-15927_Alarm 127 Low Venous Pressure Triggers Post Treatment
Add inputs.

Add inputs.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Why does this function return/set state? It is not a state handler function.

Why does this function return/set state? It is not a state handler function.

Call these blocked functions once and save values to local vars. Use local vars throughout function so you don't have to keep calling the functions. Also, why are we looking at blood recirc at all...

Call these blocked functions once and save values to local vars. Use local vars throughout function so you don't have to keep calling the functions.

Also, why are we looking at blood recirc at all? BP is never on in recirc states so no need to manage blocking blood recirculation.

Should not go to stopped state - let state machine manage state transitions. Just stop DPi and trimmer heater.

Should not go to stopped state - let state machine manage state transitions. Just stop DPi and trimmer heater.