Merge remote-tracking branch 'origin/staging' into staging
DEN-14706-can-messge-not-acked-fix
+ 143 more
- DEN-15792-hd-software-design-description
- DEN-15953-dialysate-temp-alarm-triggered-after-dialysate-recirculation-stopped
- DEN-15953-dialysate-temp-alarm-triggered-after-dialysate-recirculation-stopped-2
- DEN-15972-ui-testing-pre-treatment-disposables-check
- DEN-16090-2-sys-saline-bubbles-enter-pbo-line-during-cartridge-priming
- DEN-16090-sys-saline-bubbles-enter-pbo-line-during-cartridge-priming
- DEN-16164-fw-pre-treatment-pressure-test-failures-vent-to-saline-bag
- DEN-16200-2-syringe-pump-stuck-in-priming-state
- DEN-16200-syringe-pump-stuck-in-priming-state
- DEN-16322-arterial-pressure-window-limits-are-inaccurate-after-stabilization-period-triggers
- DEN-16325-alarm-137-hd-excessive-tilt-alarm-at-power-up
- DEN-16353-recirculation-not-available-after-rinseback-during-treatment
- DEN-16362-defect-failure-in-vtchd-740-srshd-1070
- DEN-16365-high-venous-alarm-does-not-appear-while-in-recirculation-mode
- DEN-16370-add-uf-set-rate-to-the-broadcast
- DEN-16372-srsdg-907-not-met
- DEN-16385-treatment-log-event-ids-for-heparin-pause-are-not-notated-correctly
- DEN-16391-3-fw-dialysate-temperature-alarm-not-sensitive-enough-use-tdi-moving-avg
- DEN-16394-treatment-log-for-uf-is-notated-incorrectly
- DEN-16397-after-treatment-duration-is-changed-dpo-speeds-up-and-alarm-127-venous-pressure-low
- DEN-16408-dialysate-flow-performance-decrease-at-600qd
- DEN-16428-need-a-way-or-approach-to-validate-the-srshd-1426
- DEN-16442-reservoir-fill-change-for-clearance-performance-improvement
- DEN-16446-dg-dev-test-clean-up
- DEN-16456-dpo-does-not-home-in-startup-or-end-treatment
- DEN-16487-dialin-config-helper-commands-are-not-supported-with-the-latest-hd-and-dg-builds
- DEN-16499-fw-message-list-testing-issues
- DEN-16499-fw-message-list-testing-issues-2
- DEN-16512-2-dpo-stops-during-treatment-after-reconfirming-dialysate-flow-rate
- DEN-16512-dpo-stops-during-treatment-after-reconfirming-dialysate-flow-rate
- DEN-16518-2-alarm-41-triggered-on-dvt-014-at-the-beginning-of-treatment
- DEN-16518-alarm-41-triggered-on-dvt-014-at-the-beginning-of-treatment
- DEN-16519-2-pressure-window-adjustments-do-not-reset-after-new-treatment
- DEN-16519-pressure-window-adjustments-do-not-reset-after-new-treatment
- DEN-16563-venous-pressure-readings-were-below-20mmhg-for-1.5-minutes-before-low-venous
- DEN-16563_2-venous-pressure-readings-were-below-20mmhg-for-1.5-minutes-before-low-venous
- DEN-16565-sw-updater-hd-dg-app-side-intake-from-sunrise
- DEN-16594-2-xx-blank-screen-after-selecting-end-treatment-from-alarm-141-venous-air-detected
- DEN-16594-xx-blank-screen-after-selecting-end-treatment-from-alarm-141-venous-air-detected
- DEN-16595-2-vectorcast-unit-and-integration-testing
- DEN-16595-4-vectorcast-unit-and-integration-testing
- DEN-16648-2-update-syringe-pump-positioning-in-software-per-force-sensor-button-update
- DEN-16648-update-syringe-pump-positioning-in-software-per-force-sensor-button-update
- DEN-16666-log-passing-results-of-self-test-2
- DEN-16666-log-passing-results-of-self-test-to-ui-log
- DEN-16672-2-air-trap-level-control-changes
- DEN-16672-air-trap-level-control-changes
- DEN-16680-p1b-inlet-water-shutdown-during-heat-disinfect-staging
- DEN-16691-2-limit-pressure-compensation-fo
- DEN-16691-limit-pressure-compensation-fo
- DEN-16693-institutional-settings-implementation
- DEN-16693-institutional-settings-implementation-2
- DEN-16773-add-chemical-disinfect-enable-institutional-record-2
- DEN-16778-system-does-not-check-that-air
- DEN-16779-address-cpp-check-warning
- DEN-16779-address-cpp-check-warning-2
- DEN-16782-unable-to-clear-communication-error-alarm-65-during-power-loss-2
- DEN-16819-In-Treatment-Parameter-mode-no-rejection-seen-when-hep_rate-or-hep_bol-value-is-set-to-less-than-0-p1-staging
- DEN-16838-2-uf-displays-incorrect-reject-m
- DEN-16842-After-device-reset-no-option-of-Chemical-Disinfect-and-Flush-in-cleaning-mode-with-chem-disinfect-is-enabled-in-institutional-records-p1-staging
- DEN-16857-update-alarm-text-reject-reason-for-2-hour-flush-requirement-p1-staging
- DEN-16949-low-venous-pressure-alarm-did-
- DEN-17053-release-17-firmware-update
- DEN-17053-release-17-firmware-update-usage-update
- DEN-17063-cartridge-testing-special-s-w-
- DEN-17084-release-17-prs-769-dg-device-inter-patient-disinfection-policy-is-not-satisfied-staging-p1b
- DEN-17093-blood-leak-zeroing-in-treatment-p1b-staging
- DEN-17156-blood-prime-ramp-time--45s-instead-of-55s-p1b-staging
- DEN-17156-blood-prime-ramp-time-45s-instead-of-55s
- DEN-17208-blood-leak-alarm-126-129-triggered-when-VDI-and-VDO-are-in-position-C-staging
- DEN-17213-release_20_vectorcast-staging
- DEN-17220-release_20_p1b_staging
- DEN-17271-saline-bolus-doxygen-update
- DEN-17273-release_21_cpp_check_fix-P1B-STAGING
- DEN-17273-release_21_cpp_check_fix_staging
- DEN-17299-release_22_p1b_staging
- DENBUG-101-dpi-does-not-run-when-blood-leak-sensor-fails-to-zero
- DENBUG-104-2-alarm-38-hd-fault-dpi-rotor-t
- DENBUG-104-alarm-38-hd-fault-dpi-rotor-t
- DENBUG-116-dg-allows-update-of-calibration-service-system-records-in-other-modes-besides
- DENBUG-157-2alarm-48-uf-accuracy-triggers-immediately-after-alarm-272-hd-fan-rpm-out-of-range
- DENBUG-157-alarm-48-uf-accuracy-triggers-immediately-after-alarm-272-hd-fan-rpm-out-of-range
- DENBUG-158-2-alarm-128-venous-air-bubble-triggered-again-during-post-tx-sequence
- DENBUG-158-alarm-128-venous-air-bubble-triggered-again-during-post-tx-sequence
- DENBUG-162-2-low-venous-pressure-alarm-triggers-during-normal-treatment-operation
- DENBUG-162-low-venous-pressure-alarm-triggers-during-normal-treatment-operation
- DENBUG-163-2-high-venous-pressure-alarm-130-triggers-during-normal-treatment-operation
- DENBUG-163-high-venous-pressure-alarm-130-triggers-during-normal-treatment-operation
- DENBUG-186-2-alarm-68-dialysate-gen-recirculation-error-clears-alarm-147-dialysate-gen
- DENBUG-186-alarm-68-dialysate-gen-recirculation-error-clears-alarm-147-dialysate-gen
- DENBUG-198-2-alarm-130-142-triggered-and-the-end-of-a-rinseback-end-of-treatment
- DENBUG-198-alarm-130-142-triggered-and-the-end-of-a-rinseback-end-of-treatment
- DENBUG-200-2-alarm-58-triggers-when-cartri
- DENBUG-200-alarm-58-triggers-when-cartri
- DENBUG-206-2-venous-pressure-readings-were
- DENBUG-206-venous-pressure-readings-were
- DENBUG-216-2-alarm-89-alarm_id_dg_outlet_u
- DENBUG-216-alarm-89-alarm_id_dg_outlet_u
- DENBUG-232-unable-to-clear-communication
- DENBUG-239-2-alarm-160-blood-pump-rotor-sp
- DENBUG-239-alarm-160-blood-pump-rotor-sp
- DENBUG-240-after-a-manual-rinseback-and-
- DENBUG-241-2-dialysate-flow-rate-cannot-be
- DENBUG-241-dialysate-flow-rate-cannot-be
- DENBUG-242-2-continuous-alarm-278-reservoi
- DENBUG-242-continuous-alarm-278-reservoi
- DENBUG-249-pressure-window-starting-values-p1-staging
- DENBUG-259-low-venous-pressure-alarm-did
- DENBUG-285-alarm-141-air-trap-fill-timeout
- DENBUG-285-alarm-141-air-trap-fill-timeout-end-tx-back-p1b-staging
- DENBUG-285-alarm-141-air-trap-fill-timeout-end-tx-back-staging
- DENBUG-285-alarm-141-air-trap-fill-timeout-p1b-staging
- DENBUG-290-blood-leak-alarm-during-rinseback-staging
- DENBUG-298-user-unable-to-deliver-saline-P1B-STAGING
- DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating
- DENBUG-318-rejection-ratio-setting-feature-implementation-in-manufacturing-configuration-for-treatment-staging
- DENBUG-326-syringe-switch-detection-failure_staging
- DENBUG-328-venous-pressure-alarm-did-not-trigger-P1B-STAGING
- DENBUG-328-venous-pressure-alarm-did-not-trigger-alarm-215-STAGING
- DENBUG-328-venous-pressure-alarm-did-not-trigger-when-pressure-dropped-to-0-mmHg-STAGING
- DENBUG-78-dvt-21-alarm-110-triggers-at-start-of-treatment-with-heparin-on
- DENBUG-83-alarm-278-triggered-active-res-weight-out-of-range
- DENBUG-84-heparin-bolus-does-not-resume-after-treatment-pause
- DENBUG-85-treatment-log-missing-event-for-heparin-paused-by-user-request-during-treatment
- DENBUG-96-treatment-log-inaccuracies
- DIAL-276-dg-event-event-id-15-prints-duplicate-events-with-different-state-time
- DIAL-278-fan-pwm-override
- HD-DEN-16314-alarm-206-triggers-after-resuming-from-incoming-water-conductivity-high-or-low
- STT-1965-srshd-1395-the-hd-software-shall-trigger-alarm-98-if-an-error-is-detected-on-the-battery
- STT-2824-vtchd-504-verify-alarm-is-set-point-not-set-after-3-trials
- STT-2824-vtchd-504-verify-alarm-is-set-point-not-set-after-3-trials-2
- STT-3088-bug-no-alarm-if-ui-is-not-up-for-infinity
- bugfix/DEN-16391-fw-dialysate-temperature-alarm-not-sensitive-enough-use-tdi-moving-avg
- develop
- intertek_branch_from_release_10
- master
- release/p1
- release/p1_develop
- release/p1_staging
- release/p1b/master
- release/p1b/staging
- staging
- swupdateHD
![]() |