DEN-13712 Updated function header per review comment.

DEN-13712 Fix treatment time save to look at lastMode on switch to newMode.

DEN-13712 Force Standby to disinfect sub mode after exiting treatment.

DEN-13712 Handle disinfect command from UI Setup properly (0x9A)

    • -8
    • +10
    ./App/Services/SystemCommMessages.c
DEN-13712 Added Alarms for UI to show Disinfect modes.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 18 more files in changeset.
DEN-12931 addressed the code review comment

DEN-12931 addressed the code review comments

DEN-12931 removed and old message

    • -32
    • +0
    ./App/Services/SystemCommMessages.c
DEN-12931 addressed the code review comments

DEN-12931 added isnan math function to prevent NaN values to be used in HD

DEN-12931: Updated blood flow estimation parameters for new tubing.

Merged staging and resolved the conflicts

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-12847 removed AC POST

DEN-12931: Removed POST test for checking if booting from battery.

DEN-12931: Changed syringe pump calibration to expected DAC voltage. Added verification of calibration.

DEN-12931: Refactored syringe detection alarm mgmt.

DEN-12847 addressed code review comment

DEN-12847 merged staging and addressed VectorCAST issues

DEN-12847 updated the code for VectorCAST

DEN-12931: Refactored door and pump track alarm management.

DEN-12847 addressed the code review comments and PiSquare's VectorCAST bugs

DEN-12931: Added request for UI POST final result to ensure we get result before its needed in UI POST test.

    • -3
    • +29
    ./App/Services/SystemCommMessages.c
DEN-12847 addressed the VectorCAST issues reported by PiSquare

DEN-12931: Fixed bug in setup for dialyzer prime step.

DEN-12847 Remove debug output.

DEN-12847 code review comment refactor of drift check.