Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14007: Addressed code review comments.

DEN-14007: Addressed code review comments.

DEN-14007: Fixed alarm 36 error from updates.

DEN-14007: misc fixes to homing speed updates.

DEN-14007: Tentative fix for DEN-13688 for pump homing consistency - set all pumps to same speed while homing.

DEN-14007: Increased post-tx de-prime time from 20 to 30 seconds to allow dialysate line to fully de-prime.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'staging' into DEN-14099-bubbles-in-dialysate-out-line-from-dg-to-cartridge

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14007 merged staging and resolved the merge conflicts

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DIAL-219: Fixed bug in broadcast user alarm button blockers.

DEN-14099 Change PURGE_AIR_TIME_OUT_COUNT to 4min. Remove debug output.

DEN-14099 Bubble clear mods to state machine.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-14307: partial fix to DEN-14086.

DEN-14307 addressed a code review comment

DEN-14307 addressed the code review comments

DEN-14307: Removed obsolete build switches.

DEN-14307 addressed code review comments

DEN-14007 Code Review comments update 2.

DEN-14007 Code Review comments update 1.

DEN-14099 Code review comments: code cleanup from moving bubble clear pressurize.

DEN-14307: Fixed polarity issue with reported state based alarm button blockage. Renamed battery capacity remaining variable to correct units. Stop disabling battery comm on build switch.

DEN-14307 updated the code to get the heat disinfect temperature sensors (THd) and checks it value prior to starting a treatment

DEN-14307: HD rotor vs. motor check - changed RPM threshold from 5 to 2 RPM.