DEN-8886: Rolled back last CAN clock change.

  1. … 6 more files in changeset.
DEN-8886: Changed CAN clock source to internal oscillator (16 MHz). Configured CAN1 to have 16 tQ instead of 8.

  1. … 4 more files in changeset.
DEN-8886: Updated pump control.

  1. … 7 more files in changeset.
DEN-8886: Updated CAN driver to move sample point from 75% to 87.5% to match other nodes.

  1. … 8 more files in changeset.
DEN-8886: Changed master clock to 208 MHz for better CAN clock accuracy.

  1. … 15 more files in changeset.
DEN-8886: Updates to CAN - remove debug code.

  1. … 3 more files in changeset.
Misc UF fixes. Changed CAN from FIQ to IRQ to see if helps with intermittent lockup.

  1. … 10 more files in changeset.
Removed DialOutFlow module - Leo added to separate branch. Fixed alarm lamp POST.

  1. … 18 more files in changeset.
Updated CAN interrupts config. Added a watchdog expired alarm. Added PWM duty cycle to blood flow broadcast.

  1. … 18 more files in changeset.
Increased delay between CAN bus recovery attempts. Added a little handling for UART frame/overrun errors. Updated UT. +review HD-DEN-759-1

  1. … 5 more files in changeset.
Added blood pump control per flow. Added timeout to power off request. Added more interrupt handling. Added parity checking for CAN RAM. +review HD-DEN_759-1

  1. … 12 more files in changeset.
Added dialin support for AlarmMgmt module. Refactored PC interface from UART to CAN. Added two CAN channels for PC interface. +review HD-DEN-759-1

  1. … 8 more files in changeset.
Transferred the hd_sprint5 firmware to the firmware folder

  1. … 397 more files in changeset.
Transferred the master branch firmware to the firmware folder. Added the VectorCAST folder to the repository

  1. … 397 more files in changeset.