firmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15105: Prevent DG faults from being cleared by HD.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
DEN-15164 changed the inlet water conditions persistenet alarm to 30 seconds

    • -4
    • +3
    ./App/Controllers/ConductivitySensors.c
    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
DEN-15164 removed the code to close VPi in filter flush mode when the inlet water condition is out of range

DEN-15077-fmp-flow-vs-load-cell-comparison-to-use-load-cells-as-actual-value

converted error relative to integrated flow, to error relative to load cell, as the load cell mass is a more accurate measurement.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -1
    • +1
    ./App/Controllers/TemperatureSensors.c
DEN-14999-2: Removed unused #define.

DEN-14999 updated the code based on the alarms table

    • -9
    • +9
    ./App/Controllers/ConductivitySensors.c
    • -6
    • +6
    ./App/Controllers/TemperatureSensors.c
DEN-14999-2: Addressed code review comments.

DEN-14999-2: Exempt fpga comm errors for 1st second after power up.

DEN-14999, DEN-15088 added persistent alarm to the DG fluid leak

DEN-14999 added 5-second delays prior to collecting acid and bicarb conductivity values during acid and bicarb priming

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    ./App/Controllers/ConductivitySensors.c
    • -2
    • +2
    ./App/Controllers/TemperatureSensors.c
    • -1
    • +1
    ./App/Modes/ModeChemicalDisinfectFlush.c
    • -1
    • +1
    ./App/Modes/ModeHeatDisinfectActiveCool.c
  1. … 3 more files in changeset.
DEN-14999: merged staging

    • -25
    • +49
    ./App/Controllers/ConductivitySensors.c
    • -26
    • +21
    ./App/Controllers/TemperatureSensors.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    ./App/Controllers/ConductivitySensors.c
    • -2
    • +2
    ./App/Controllers/TemperatureSensors.c
Merge branch 'staging' into DEN-14646-fpga-alarm-properties-and-timing-3

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14646 Correction per code review comments

    • -3
    • +2
    ./App/Controllers/ConductivitySensors.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DG-DEN-15016_System Faults During Treatment Then Returns TO The Standby Screen

Update for code review.

DIAL-222: changed set heater duty cycle to consider overriden values on reset.