DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
Merge branch 'staging' into DEN-15105-dialyzer-flip

    • -254
    • +251
    ./App/Controllers/Voltages.c
DEN-15105: re-enabled a/c power loss alarm.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15105: Fixed dry s.t. countdown going past 0 issue.

DEN-15105: Fixed DEN-14309.

DEN-15105: Implemented pre-tx patient connection states to handle dialyzer invert.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15105: Started adding pre-tx patient connection sub-states.

DEN-15049 Code review changes to accomodate using CPLD for power loss detection

DEN-15049 Comment correction / clarification

DEN-15105: rolled back req for DG POST to pass to start treatment workflow for now - not working.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15105: Fix for DEN-15154 - saline bolus issue.

DEN-15105: Resolved some compiler warnings. Require POST to pass before starting treatment workflow.

DEN-15049 Correct header

DEN15049 Correct header

Merge branch 'staging' into DEN-15049-fw-hd-voltage-not-trigger-alarm-76-when-voltage-out-of-range-in-line

# Conflicts:

# firmware/App/Controllers/Voltages.c

    • -235
    • +244
    ./App/Controllers/Voltages.c
DEN-15105: Addressed code review comment.

DEN-15105: Updated blood tubing volumes for blood prime and rinseback.

DEN-15105: Skip reservoir draining in post-tx if DG is in fault.

DEN-15049 Correct voltage monitoring across all voltages.

DEN-15105: Fix for DEN-15116.

DEN-15105: Removed alarm 227. Fixed bug DEN-15102.

    • -20
    • +21
    ./App/Modes/ModeTreatmentParams.c
DEN-15110 Use 800ml/min reservoir fill rate for all dialysate flow rates.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo