ModeTreatment.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-15267: Refactored bubble detector driver to allow modes to push enable/disable instead of determining based on current op mode. Added support for new Dialin command to enable/disable venous bubble alarm detection.

  1. … 11 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
DEN-15183: Addressed code review comments. Fixed DEN-15194.

  1. … 1 more file in changeset.
DEN-15183: indicate treatment over in treatment times broadcast if alarm preventing alarm resumption.

DEN-15183: Addressed DEN-15191 by zeroing treatment time remaining in treatment times broadcast msg if an alarm has blocked resumption of treatment.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 21 more files in changeset.
DEN-15105: Skip reservoir draining in post-tx if DG is in fault.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 30 more files in changeset.
DEN-14999 updated the alarms table

  1. … 20 more files in changeset.
DEN-14999-2: prevent treatment resumption from rinseback/recirc states if an alarm that blocks resume has been triggered previously in the treatment.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
DEN-14906: fixed merge conflicts from staging.

  1. … 11 more files in changeset.
DEN-14906: Fixed isues that arose when disabling re-prime feature.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
DEN-14906: Restored 30 minute interval for dialyzer re-prime. Reduced re-prime time from 40 to 35 seconds. Fixed so treatment time is paused during dialyzer re-prime.

  1. … 1 more file in changeset.
DEN-14906: Added dialyzer reprime state to treatment mode states broadcast message.

  1. … 3 more files in changeset.
Merge branch 'staging' into DEN-14689-messaging-bugs

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-14888: Increased minimum UF rate from zero to 100 mL/hour.

  1. … 3 more files in changeset.
DEN-14689: message naming sync

  1. … 11 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Fixed merge conflicts with local merge from staging.

  1. … 5 more files in changeset.
DEN-14561 Code review comment update.

DEN-14561: Set critical data values for UF and Tx duration settings when changed in mid-treatment.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-14749: Reset preload flag in standby mode instead of after retract/seek/prime operations. Increase alarm threshold for FMD vs. LC check in wet s.t. from 5% to 10%. Refactor alarm audio and 1 minute min tx time config flags. Removed test code in reservoir code.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 46 more files in changeset.
DEN-14549: Update to arterial/venous pressure limit alarm logic. Update measured UF volume in all dialysis states. Remove dialysis and UF start states.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 112 more files in changeset.