Merge branch 'staging' into DEN-15245-heparin-empty-syringe-alarm-notification-not-displayed-to-user-if-another-alarm-is

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-15245: Addressed DEN-14818.

DEN-15199 addressed the code review comments

DEN-15199 addressed the code review comment

DEN-15199 merged staging

  1. … 3 more files in changeset.
DEN-15199 renamed the test configuration enums

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-15267: Refactored bubble detector driver to allow modes to push enable/disable instead of determining based on current op mode. Added support for new Dialin command to enable/disable venous bubble alarm detection.

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
Merge branch 'staging' into DEN-14262-vba-homing-fails-when-cartridge-is-loaded-on-device

DEN-15199 work in progress continued implementing the test configurations code

  1. … 5 more files in changeset.
DEN-14262: Addressed code review comments.

DEN-14262: Force re-home in no cartridge self-tests state. Addressed DEN-14171.

  1. … 3 more files in changeset.
DEN-14262: Addressed DEN-15173 re: getting user confirmation on end treatment requests in rinseback, recirculate and end treatment states.

DEN-15199 added test configuration code

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15203: code review comments.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15214: Removed alarm 47 - UF rate alarm.

DEN-15023: fixed max heparin volume check.

DEN-15023: reworded defines to handle advised fill volume and hw tolerance. removed positive position for force sensor.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15183-2: Fixed issues with reconnect state of treatment recirculation.

DEN-15183-2: Fix init of tx recirc to correct state.

DEN-15183-2: Estimate initial DPo rate correction offset based on data collected by Systems team to improve UF control at treatment start.

  1. … 2 more files in changeset.
DEN-15183-2: Minor fixes to Tx recirculation updates.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-15183: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-15023: reworked syringe empty check to consider force and hw variance. Added t_par check for maximum heparin. renamed syringe occlusion and gave occlusion alarm persistence outside of priming