DEN-15282: Changed rinseback volume from 400 to 350 mL per Systems/Clinical input.

DEN-15282: Allow 2 minutes between UI comm check-ins while in service mode (due to lockdown possibility).

    • -3
    • +10
    /firmware/App/Services/SystemComm.c
DEN-15282: Reverted pressure stabilization time back to 60s from 30s. During rinseback run and paused states, allow resume and end tx options for alarms - resume will resume rinseback.

    • -2
    • +18
    /firmware/App/Modes/ModeTreatment.c
DEN-15282: increased timeout for rinseback saline bolus to 25 sec from 20 sec.

DEN-15282: Fixed rinseback bolus rate sent to UI for display.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-15510 to staging.

Merged staging to DEN-15510.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
DEN-15510: Fixed RB volume at 400 mL. Changed RB bolus volume from 10 to 50 mL.

DENBUG-63 Updated coefficients for DPi flow calculations.

    • -6
    • +6
    /firmware/App/Controllers/DialInFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15656 to staging.

Merged staging to DEN-15656.

DEN-15656: publish 200 rate for rinseback additional.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15330-dialysate-temp-alarm-recovery-takes-a-long-time-no-reservoir-switching' into staging

DEN-15656: Addressed code review comment.

DEN-15656: Addressed DEN-15654 issue where pressure limits did not go to stabilization limits on transition to treatment paused state.

DEN-15656: Addressed DEN-15656 issue where pressure stabilization would not occur on alarm resume if blood was recirculating during alarm. Addressed DEN-15652 where rinseback saline bolus was delivered at 50 mL/min instead of 200 mL/min.

DEN-15330: restored conditional formatting

DEN-15330: fixed formatting.

    • -16
    • +14
    /firmware/App/Services/Reservoirs.c
Merge branch 'staging' into DEN-15330-dialysate-temp-alarm-recovery-takes-a-long-time-no-reservoir-switching

DEN-15330: Combined switch reservoir conditionals.

    • -20
    • +2
    /firmware/App/Services/Reservoirs.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
    • -0
    • +16
    /firmware/App/Services/Reservoirs.h
Merged DEN-15489