DEN-15614 updated the chemical disinfect VPd to drain in the last drains, DEN-16354 changed the heat disinfect 500 mL timeout to 2 minutes to align with the reqruirements

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-15614 updated the chemical disinfect VPd to drain in the last drains

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.h
Merged DEN-16309

DEN-16309 merged staging

    • -6
    • +107
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -2
    • +6
    /firmware/App/Modes/ModeHeatDisinfect.h
DEN-16309 addressed a code review comment

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16093 changes merge in to staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.h
Merge branch 'DIAL-281-requesting-heat-disinfect-timer-overrides' into staging

DEN-16323, DEN-16277 updated the chemical disinfect to get the level of the reservoir under disinfection more accurately

    • -20
    • +59
    /firmware/App/Modes/ModeChemicalDisinfect.c
Merge branch 'staging' into DEN-16309-heat-disinfect-has-spikes-in-temperature-near-95c-which-will-cause-water-to-boil-at

DEN-16309 updated the code. Changed the load cells' full level from 1900 to 1850 mL in heat disinfect to be consistent with the rest of the cleaning modes

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DIAL-281: Code review comments.

    • -3
    • +3
    /firmware/App/Modes/ModeHeatDisinfect.c
DIAL-278: implementation of heat disinfection state timer overrides.

    • -4
    • +105
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -0
    • +4
    /firmware/App/Modes/ModeHeatDisinfect.h
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DIAL-276-dg-event-event-id-15-prints-duplicate-events-with-different-state-time' into staging

DEN-16309 added a new heat disinfect state

    • -7
    • +11
    /firmware/App/Controllers/Heaters.c
    • -10
    • +62
    /firmware/App/Modes/ModeHeatDisinfect.c
    • -0
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.h
DEN-16309 removed the clearance period some of the persistence alarms

DEN-16093 fixed code review comments

Merge branch 'staging' into DIAL-276-dg-event-event-id-15-prints-duplicate-events-with-different-state-time

DEN-16093-Increased conductivity percentage check and time out values

take latest changes from staging

    • -4
    • +58
    /firmware/App/Services/SystemComm.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces