dgfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15702 Code review update - remove spaces.

DEN-15702 Indent for code review comments.

    • -44
    • +44
    /firmware/App/Controllers/Pressures.c
DEN-15702 Check VPI Open for Inlet Conductivity, Pressure, temperature. If VPI Closed, clear alarms.

    • -8
    • +17
    /firmware/App/Controllers/Pressures.c
    • -14
    • +11
    /firmware/App/Modes/ModeStandby.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c
Merge branch 'DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it' into staging

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Corrected comment alignment

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Undated 5second timeout to 10 seconds

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15683-fw-alarm-handling-of-concentrate-pumps-is-incorrect' into staging

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15701 to staging.

DEN-15701: handle stop DG command from HD while in standby water sampling states.

    • -15
    • +16
    /firmware/App/Modes/ModeStandby.h
DEN-15683 Correct alignmen

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15647

DEN-15647 addressed the code review comments

DEN-15683 Updates per code review comments

DEN-15683 Correct proper alarming of concentrate pumps

DEN-15610 fixed DG flush to write basic flush complete timestamp in the usage record

DEN-15647 updated the heat disinfect active cool mode to deenergize the actuators prior to transitioning to cancel water path state

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15571

DEN-15571 updated the code

DEN-15571, DEN-15398 changed the fill mode full speed prime time from 8 seconds to 7 seconds

DEN-15571 addressed the code review comments

    • -14
    • +11
    /firmware/App/Modes/ModeChemicalDisinfect.c