Typo.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Blood pump only stops for user stop.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Remove redundant condition.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Simplified. Addressed CR comments.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Add restart flag.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Control update for Christina's observation

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Updates to address comments.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Corrected comments and alignment

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

HW update only on flag changes

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Minimal change to do "all stop" for user stop button in treatment recirculation.

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Reversed working revisions, to try minimal change.

  1. … 3 more files in changeset.
DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2

Update to user stop state, and hard stop functions.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15804 reverted the saline bolus changes

DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

cleared alarm 120 in self test

DEN-15893-cartridge-lever-is-open-alarm-when-cartridge-lever-is-closed

Cleared alarm 120, prior to initiating alarm 85

Required latch closed before allowing alarm 85

Paused in WET self test until bubbles clear or latch closed.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15664-blood-pump-stop-time-exceeded-alarms-124-125-165-and-212-alarm-timer-feature-is' into staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into staging

DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus

Stop the pumps more forcefully, open loop, prior to switch valves for the saline bolus.

DEN-15664 Remove treatment end publish overrides.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-15536 addressed the code review comments

Merge branch 'staging' into DEN-15536-fw-usability-dry-self-tests-did-not-repeat-after-dry-self-test-failure

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-15536: Addressed DEN-15898.