DEN-15802 Correct alarm 191.192 testing based upon specific modes.

STRL-15802 Add Heat Disinfect Active Cool to logic to test for 191, 192

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Corrected comment alignment

DEN-15691-heat-disinfect-reservoir-leak-alarm-triggers-after-5s-when-srsdg-512-and-521-say-it

Undated 5second timeout to 10 seconds

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15701: handle stop DG command from HD while in standby water sampling states.

DEN-15683 Correct alignmen

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15647 addressed the code review comments

DEN-15683 Updates per code review comments

DEN-15683 Correct proper alarming of concentrate pumps

    • -20
    • +56
    ./Controllers/ConcentratePumps.c
DEN-15610 fixed DG flush to write basic flush complete timestamp in the usage record

DEN-15647 updated the heat disinfect active cool mode to deenergize the actuators prior to transitioning to cancel water path state

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15571 updated the code

DEN-15571, DEN-15398 changed the fill mode full speed prime time from 8 seconds to 7 seconds

DEN-15571 addressed the code review comments

DEN-15571 updated the chemical disinfect drain status function

DEN-15443 added code to transition back to ramp state in concenrate pumps if the new requested target speed is greater than the threshold

DEN-15571 made changes in the chemical disinfect mode to not trigger the max temeprature and conductivity out of range until the end of the chemical disinfect mode

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
Merge branch 'staging' into DEN-15367-hd_dg_test_configuration_setup