DEN-16725 converted the calculated UF volume from mL to L prior to checking it against the lower and upper range of the UF volume treatment parameters in the set duration time function in the mode treatment
DEN-16728 made a change in the treatment parameters to be able to use the 1-minute treatment time. DEN-16725 made a change in the treatment duration function to check the edges of the ultrafiltration volume