DEN-9906 updated the code to support events

  1. … 4 more files in changeset.
DEN-9906: Fixed merge conflicts from staging.

  1. … 47 more files in changeset.
DEN-9906: Added new volume checks to saline bolus, rinseback and blood prime deliveries.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-9480 addressed the code review comments

  1. … 1 more file in changeset.
DEN-9480 addressed the code review comments

  1. … 3 more files in changeset.
DEN-9906: Fixed misc TBDs in comments.

  1. … 5 more files in changeset.
DEN-9906 updated the drain command with start/stop

  1. … 2 more files in changeset.
DEN-9906: Increased upper limit of arterial pressure limit per PRS change.

Fixed bugs DEN-10983 and DEN-10982. Removed obsolete code associated with obsoleted components/features.

  1. … 7 more files in changeset.
DEN-9906: Fixed DEN-10792.

DEN-9906: Fixes for DEN-9931, DEN-10155 and DEN-10974.

DEN-9906: Added overrides for blood pump rotor count and arterial pressure offset. Removed unused handling of DG data and related functions/variables. Fixed DEN-10973.

  1. … 10 more files in changeset.
DEN-9906: Fixed bugs DEN-10792 and DEN-10803.

DEN-9906: Increased UI wait time to max 40 seconds. Misc changes to message buffers.

  1. … 3 more files in changeset.
DEN-10158 addressed the HD valves by holding the next command

  1. … 7 more files in changeset.
Run DPi open loop too for clearance testing.

  1. … 2 more files in changeset.
No inline pressure checks if disabled by build flag. If build flag says no prime, fill reservoirs all the way.

  1. … 2 more files in changeset.
DEN-9906: Removed dialysate pump occlusion sensors. Started changes to replace FMB with flow estimation. Added test build switch to run BP open loop during treatment.

  1. … 9 more files in changeset.
DEN-9906: Fixed logic preventing low venous alarm when venting air trap. Updated inline pressure alarm limit min/max levels.

  1. … 1 more file in changeset.
DEN-9906: Replaced local statics ffor blood and dialysate rates in treatment mode to calls via critical datas for those rates. Fixes problem where rate change did not update fill volume for DG reservoirs.

DEN-9906: Fixed general event sub-mode change issue.

DEN-9906: Added f/w support for Dialin cmd to set syringe pump DAC ref voltage.

  1. … 3 more files in changeset.
DEN-9906: Optimized HD broadcast messaging.

  1. … 23 more files in changeset.
DEN-9906: Addressed DEN-8719. Finished sending sub-mode change events. Minor changes per DEN-10202.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 97 more files in changeset.
DEN-9906: For consistency, made all rinseback and recirculate timeout warnings 2 minutes before alarms. Fixed bug DEN-10168. Do not trigger blood sitting alarms from treatment stop sub-mode if rinseback completed (no blood in line). Set rinseback completed flag if an additional rinseback volume puts us over 300 mL.

DEN-9906: Trigger appropriate alarms when blood is sitting too long in treatment end mode DEN-10165.

DEN-9906: Fixed rinseback base countdown time after rinseback completed (DEN-10162).

DEN-9906: Fixed bug DEN-10162. Ignore low venous pressure while venting air trap.

  1. … 2 more files in changeset.