dgfirmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-334 reverted the heat disinfect reservoir transfter target flow back 0.5 L/min

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
updated the code to initialize the variables in between the treatments without a need to power cycle the unit

    • -1
    • +1
    /firmware/App/Controllers/Thermistors.c
    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-327

DENBUG-327 updated the VectorCAST test cases

DENBUG-327 updated the baro sensor mfg CRC check

DENBUG-327 added code to make the baro sensor MFG CRC as LSB

DENBUG-327 updated the baro sensor code

    • -11
    • +25
    /firmware/App/Controllers/Pressures.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-17298

DEN-17298 updated the VectorCAST test cases

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-318

DENBUG-318 addressed the code review comment

DENBUG-318 addressed the code review comments

DENBUG-318 updated the VectorCAST test cases

DENBUG-318 added code to request the DG institutional values in mode standby until they are received

    • -12
    • +26
    /firmware/App/Modes/ModeStandby.c
DENBUG-318 added code to keep requesting DG institutional values once start DG has been sent until the values are received

DENBUG-318 changed the min inlet water conductivity to recovery offset to 10 uS/cm. Updated the vectorcast test cases

DENBUG-318 added the request command to get the DG institutional record from HD upon starting the DG

DENBUG-318 added an RO rejection ratio event

DENBUG-318 changed the max RO rejection ratio to min RO rejection ratio in percent and updated the mode fill calculations to check the RR

DENBUG-320 removed the acid and bicarb used volumes in mode fill broadcast

DENBUG-318 removed the inlet recovery conductivity alarm #define

DENBUG-318 changed the inlet water conductivity low alarm recovery

DENBUG-318 added a new event for min inlet water conductivity alarm from HD institutional record

DENBUG-318 updated the code to receive all the DG institutional record items. Added code to use the min inlet water conductivity alarm from the institutional record

DENBUG-319 added code to receive a command from dialing to enable the mode fill chemicals test