dgfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DIAL-243: Fixed header.

DIAL-243: fixed integrated volume data type from u32 to f32

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/FlowSensors.c
Merged DEN-13154 into staging.

DEN-13154: Added raw flow rates to DG flow broadcast messages.

    • -4
    • +6
    /firmware/App/Controllers/FlowSensors.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-14834

DEN-14834 updated the code to just send the heater command in Idle and Drain if the primary heater is not on. Addressed the code review comments

    • -10
    • +0
    /firmware/App/Services/Reservoirs.h
Merge branch 'staging' into DEN-14834-ro-flow-drops-to-zero-and-cp-measured-speed-1500-temporarily-during-pre-treatment

DEN-14834 updated that get target primary heater function to accomodate the changes from pre-treatment. Removed the range check from the drain pump measured RPM

    • -10
    • +6
    /firmware/App/Controllers/DrainPump.c
    • -15
    • +17
    /firmware/App/Controllers/Heaters.c
    • -29
    • +44
    /firmware/App/Services/Reservoirs.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14788-DIAL-222' into staging

Merge branch 'staging' into DEN-14788-DIAL-222

DEN-14788: code review comments

DEN-14834 updated the code to not check the rejection ratio and CPo if the target fill is less than 550 mL. Updated the heaters code to make sure the stop heaters command stops the heaters unless a new start command is issued

    • -31
    • +39
    /firmware/App/Controllers/Heaters.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Merged DEN-14859

DEN-14859 addressed a code review comment

DEN-14859 addressed a code review comment

    • -2
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-14859 updated the code

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-14859 updated the state timers of the cleaning modes

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
DEN-14788: DEN-14831: Fixed inlet pressures not being monitored across all modes. Fixed inlet temperature to not be monitored in flush drain state.

DEN-14788: Fixed heaters duty cycle override to immediately set PWM instead of waiting for control state machine to set PWM.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14788-sw-s91-mg-cp-persistence' into staging