DEN-16203: code review comments

DEN-16203: changed dopMeasuredRate to handle set flow rate in open loop state.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16239 added the blood leak alarm disable test configuration

DEN-16200: Addressed code review comment. Addressed DEN-16229 re: Heparin not dispensing after alarm w/ blood recirc.

  1. … 2 more files in changeset.
DEN-16200: Changed venous occlusion alarm to only trigger when in pre-tx mode per discussion with Blaine.

DEN-16200: Added timeout to Heparin prime which will trigger occlusion alarm.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
Merged staging to DEN-14815.

DEN-14815: Addressed DEN-15885. Fixed compiler warning.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-16018 added more comment for clarity. DEN-16149 Changed the default active reservoir to reservoir 1 to be matched with DG on power up

  1. … 1 more file in changeset.
DEN-14815: Switched to new CPLD based AC power loss check function.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-16166: added blood leak self test event.

DEN-15993 Make same logic change for cartridge level (AlarmID 85)

DEN-15993 Correct resume handling for front door

DEN-15927-alarm-127-low-venous-pressure-triggers-post-treatment-1

Remove unnecessary define

DEN-15927-alarm-127-low-venous-pressure-triggers-post-treatment-1

Incorporated state change, rather than flow.

DEN-15927-alarm-127-low-venous-pressure-triggers-post-treatment-1

Added a minimum flow requirement to provide a small additional margin to stabilize venous pressure.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DENBUG-62: Addressed code review comment.

DENBUG-62: Varied exemption period for low venous pressure alarm following an air trap fill based on Qb.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-15886: Added events for air trap fill and air pump on/off events.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.