Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Updated

Updated

Updated

Updated

I think this function should be in ModeTreatment or Messaging.

I think this function should be in ModeTreatment or Messaging.

update params in header.

update params in header.

you are right, I missed it here. Fixed.

you are right, I missed it here.
Fixed.

Move pop to be right after last structure to be packed.

Move pop to be right after last structure to be packed.

I thought acid fields were F32?

I thought acid fields were F32?

Update inputs/outputs after updating function.

Update inputs/outputs after updating function.

We need to look for user request to resume from pause where we would restart BP (at last rate we were at before pause) and re-open art/ven pinch valves and transition back to either ramp or run sta...

We need to look for user request to resume from pause where we would restart BP (at last rate we were at before pause) and re-open art/ven pinch valves and transition back to either ramp or run state (whichever state we were in when we paused).

In this state, we still need to signal transition to dialysis when target volume is reached (same as in ramp state). We also need to look for user request to pause (same as in ramp state).

In this state, we still need to signal transition to dialysis when target volume is reached (same as in ramp state).
We also need to look for user request to pause (same as in ramp state).

We do need an else if here for when user does change the blood flow rate. In that case, we would want to go to run state where we would run continuously at the requested rate without ramping anymor...

We do need an else if here for when user does change the blood flow rate. In that case, we would want to go to run state where we would run continuously at the requested rate without ramping anymore. Ask Nico which message UI sends when user changes blood flow rate.

And we need another else if for when user asks to pause blood prime and we would stop BP (and maybe close art/ven pinch valves) and go to pause state.

Input and Output is cumulativeBloodPrimevolume_mL.

Input and Output is cumulativeBloodPrimevolume_mL.

signal function will take us out of blood prime, so no reason to change state here. We would not want to go to run state even if we weren't leaving - run state is for when user changes blood flow r...

signal function will take us out of blood prime, so no reason to change state here. We would not want to go to run state even if we weren't leaving - run state is for when user changes blood flow rate which they have not done here.

Raghu is fixing this in his Tx Params branch, but you should fix it too. TUBING_BLOOD_PRIME_VOLUME_ML is not a treatment parameter. See Raghu's fix and duplicate it here.

Raghu is fixing this in his Tx Params branch, but you should fix it too. TUBING_BLOOD_PRIME_VOLUME_ML is not a treatment parameter. See Raghu's fix and duplicate it here.

LEAHI-TD-FIRMWARE-LDT-3126_Blood Prime - TD
LEAHI-TD-FIRMWARE-LDT-3126_Blood Prime - TD
removed both enums

removed both enums

Updated

Updated

Updated

Updated

LEAHI-TD-FIRMWARE-LDT-3559_Unable to start treatment via Dialin or UI
LEAHI-TD-FIRMWARE-LDT-3559_Unable to start treatment via Dialin or UI
LDT-3559 fixed the target blood prime vol calculation equation.

Had that question. Thanks Changed to else block

Had that question. Thanks
Changed to else block

Updated

Updated

This will be gone once we have institutional records, do you still want me to move it to a separate func?

This will be gone once we have institutional records, do you still want me to move it to a separate func?

Fixed

Fixed

Can we uncomment these and send msg?

Can we uncomment these and send msg?

Can we uncomment this and send the limits to the UI?

Can we uncomment this and send the limits to the UI?

Should we cast these to (S32)?

Should we cast these to (S32)?