dd-firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.

Missing comment.

Missing comment.

update comment. next one as well.

update comment. next one as well.

blank line after the banner and maintain alphabetical order of include files.

blank line after the banner and maintain alphabetical order of include files.

LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
LEAHI-DD-FIRMWARE-LDT-4001_HDF - FW Implementation - 2/4:
Done.

Done.

I think we also need this to be uncommented, as we were running into an alarm because of this.

I think we also need this to be uncommented, as we were running into an alarm because of this.

Added new field for d1 filtered temperature and restored the pressure sensor temperature.

Added new field for d1 filtered temperature and restored the pressure sensor temperature.

The latest code from staging has an update and fixes this comment. It's not showing here, since the file was already reviewed.

The latest code from staging has an update and fixes this comment. It's not showing here, since the file was already reviewed.

as discussed, these were outcome of the test results. we will review these gains with latest controller.

as discussed, these were outcome of the test results. we will review these gains with latest controller.

Done.

Done.

Done.

Done.

LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-DD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop si...

On DD f/w side, this should just be a simple pump controller. States should be Off and On (control to commanded Qsub). There is no flow sensor for feedback control, so assuming this is open loop similar to UF pump.
GenDialysate mode will have an HDF sub-state that signals this controller from Off to On and would also call separate controller for heater and separate (Switches) monitor for door sensors.

With all of that in mind, I think this unit should have a simpler name like SubPump.

I think there should be some state monitoring for potential alarm(s) in here too. Similar to BP door on TD f/w side where the current mode/state can signal this unit that it expects specific switch...

I think there should be some state monitoring for potential alarm(s) in here too. Similar to BP door on TD f/w side where the current mode/state can signal this unit that it expects specific switch state(s) (e.g. while in HDF state, we would expect the cap to be open and stowed and we would want to alarm if that was not the case).

Remove extra blank line.

Remove extra blank line.

Where is call to HDF controller?

Where is call to HDF controller?

Monitors should come before mode mgmt above.

Monitors should come before mode mgmt above.

What about HemoDiaFiltration.h?

What about HemoDiaFiltration.h?

What about initHDF?

What about initHDF?

Where is HemoDiaFiltration.h?

Where is HemoDiaFiltration.h?

Why does this file name have pound signs pre/post?

Why does this file name have pound signs pre/post?

This macro is already defined in Common.h.

This macro is already defined in Common.h.

Align all "=" or none of them.

Align all "=" or none of them.

DD HDF control data published should just be state and set rate. TD HDF data would have volume, etc...

DD HDF control data published should just be state and set rate. TD HDF data would have volume, etc...

This looks like pump control function. I think TD f/w should be doing this when setting Qsub before sending to DD f/w. Again, I think DD f/w should just be controlling pump to the Qsub rate given b...

This looks like pump control function. I think TD f/w should be doing this when setting Qsub before sending to DD f/w. Again, I think DD f/w should just be controlling pump to the Qsub rate given by TD f/w.
HDF volume should be calculated by TD f/w by integrating commanded Qsub over time.