Space after if.

Space after if.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

remove empty space after d66pressure variable. or match the indent to previous statement.

remove empty space after d66pressure variable. or match the indent to previous statement.

Add a TODO comment (remove later): this should be handled in IO/FP module.

Add a TODO comment (remove later): this should be handled in IO/FP module.

Update comment.

Update comment.

Add comment.

Add comment.

Depressurize?

Depressurize?

Fix the typo.

Fix the typo.

moved to general task

moved to general task

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband co...

There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband control check, use the same for 'runPIcontroller' function as well.

This should return POST conductivity.

This should return POST conductivity.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

Init PI Controls missing here.

Init PI Controls missing here.

13734.87F

13734.87F

Revise the code per new UF implementation.

Revise the code per new UF implementation.

looks like this function sets the actuators every time when the same state is set by TD?

looks like this function sets the actuators every time when the same state is set by TD?

Remove Dialysate from the handler name.

Remove Dialysate from the handler name.

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Can we move this exec at task level, independent of Drybicart operations?

Can we move this exec at task level, independent of Drybicart operations?

Comment missing?

Comment missing?

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5...

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5 cycles of balancing chamber switching time.

both acid and bicarb mix volume

both acid and bicarb mix volume

Add MULTIPLIER at the end.

Add MULTIPLIER at the end.

Initialize at init function.

Initialize at init function.