Has integrity unit been added to TD project?

Has integrity unit been added to TD project?

Where are these being triggered?

Where are these being triggered?

Is this necessary? If so, why?

Is this necessary? If so, why?

LEAHI-TD-FIRMWARE-LDT-3259_Software Update - TD
LEAHI-TD-FIRMWARE-LDT-3259_Software Update - TD
Done

Done

Done

Done

I don't think we need this.

I don't think we need this.

Status should only be TRUE if msg accepted and signal sent. Move into if statement above.

Status should only be TRUE if msg accepted and signal sent. Move into if statement above.

Done

Done

Remove blank line.

Remove blank line.

Done

Done

Done

Done

Done

Done

Done

Done

Done

Done

Use alarmStatus structure to check whether alarm resume is allowed. proprs only looks at the passed alarm.

Use alarmStatus structure to check whether alarm resume is allowed. proprs only looks at the passed alarm.

This should not be necessary. Already included by AlarmMgmt.h which includes TDCommon.h which includes Common.h which includes AlarmMgmtTD.h.

This should not be necessary. Already included by AlarmMgmt.h which includes TDCommon.h which includes Common.h which includes AlarmMgmtTD.h.

Dialysate temperature alarms should be TRUE for auto resume.

Dialysate temperature alarms should be TRUE for auto resume.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

When Systems gives me Leahi terms, they will be very small fractions.

When Systems gives me Leahi terms, they will be very small fractions.

Fixed.

Fixed.

You're right. Old comment from Denali. Fixed.

You're right. Old comment from Denali. Fixed.

Came from Denali, so I believe this is what they wanted. The nearly 5 minutes is the one the user sees.

Came from Denali, so I believe this is what they wanted. The nearly 5 minutes is the one the user sees.