We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband co...

There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband control check, use the same for 'runPIcontroller' function as well.

This should return POST conductivity.

This should return POST conductivity.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

Init PI Controls missing here.

Init PI Controls missing here.

13734.87F

13734.87F

Revise the code per new UF implementation.

Revise the code per new UF implementation.

looks like this function sets the actuators every time when the same state is set by TD?

looks like this function sets the actuators every time when the same state is set by TD?

Remove Dialysate from the handler name.

Remove Dialysate from the handler name.

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Can we move this exec at task level, independent of Drybicart operations?

Can we move this exec at task level, independent of Drybicart operations?

Comment missing?

Comment missing?

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5...

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5 cycles of balancing chamber switching time.

both acid and bicarb mix volume

both acid and bicarb mix volume

Add MULTIPLIER at the end.

Add MULTIPLIER at the end.

Initialize at init function.

Initialize at init function.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Lets remove this.

Lets remove this.

LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
Looks On event handled later. One event should be fine for On condtion.

Looks On event handled later. One event should be fine for On condtion.

Space after 0

Space after 0

prevDryBicartState

prevDryBicartState

I guess, you should use OnOff status as second argument, since isBalChamberSwitchingActive varirable set after this function, it won't have latest status.

I guess, you should use OnOff status as second argument, since isBalChamberSwitchingActive varirable set after this function, it won't have latest status.

prevBalChamberState?