Lines of code count not available
the else part here, should handle greater than Qd: 300?
Get this logic into another function and call the function here.
Is this supposed to be difference in pressure, not the spent dial pressure?
I do not see TD use case in any of the PI controller functions. so why TD for this function ?
Why do we have a local sub state variable that we don't do anything with?
This is to avoid compilation warning
Add TODO comment to restore later.
I don't think these coefficient access functions need to be public. Consider making them private.
These seem more like concentrate doses, not dialysate doses. Consider renaming.
What is this?
Add blank line after declarations.
Are these coming back? If so, add TODO comment to restore later. If not, delete them.
else TD ?
This is a structure, not an enumeration.
updated.
done.
4 bytes reg.
4 byte reg.