Space after if.

Space after if.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

Can we bring depressurize as well one of the states like fill, supply and drain? And the same way when one of the operations initiated and other should not kick in.

remove empty space after d66pressure variable. or match the indent to previous statement.

remove empty space after d66pressure variable. or match the indent to previous statement.

Add a TODO comment (remove later): this should be handled in IO/FP module.

Add a TODO comment (remove later): this should be handled in IO/FP module.

Update comment.

Update comment.

Add comment.

Add comment.

Depressurize?

Depressurize?

Fix the typo.

Fix the typo.

moved to general task

moved to general task

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
Revise the code per new UF implementation.

Revise the code per new UF implementation.

looks like this function sets the actuators every time when the same state is set by TD?

looks like this function sets the actuators every time when the same state is set by TD?

Remove Dialysate from the handler name.

Remove Dialysate from the handler name.

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Add Comment. Rename macro as UF_VOL_ML_PER_TASK_INTERVAL

Can we move this exec at task level, independent of Drybicart operations?

Can we move this exec at task level, independent of Drybicart operations?

Comment missing?

Comment missing?

LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 12 more files in changeset.
LDT-4409 drybicart and dialysate mixing firmware implementation

  1. … 1 more file in changeset.
LDT-4409 addressed review comments

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
Michaels latest code resolved this issue. make sure when you merge to staging, pull the latest from staging. Check with Michael on code commit in staging.

Michaels latest code resolved this issue. make sure when you merge to staging, pull the latest from staging. Check with Michael on code commit in staging.

LDT-4409 addressed review comments

  1. … 5 more files in changeset.