DG-DEN-16280

As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

As we discussed, I think alarm 150 and related inlet water alarms are already set to block triggering during rinseback and post-treatment modes. So these changes are likely not needed.

This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely re...

This was a specific change for this alarm (150), and the associated alarms in this routine. It appeared there were similar mode limits for temperature as well. Pressure limits were more uniquely related to valve position, I thought.

I had seen the related changes in 59, and included them in my develop build.

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature? FYI - getHDOperationMode()...

We don't trigger inlet water conductivity alarm in post-treatment mode? Any reason why we're not just saying "== MODE_POST"? What about inlet water pressure/temperature?

FYI - getHDOperationMode() was not working - HD op mode handler was mishandling message - fixed in my DENBUG-59 branch, but not yet in staging.

DG-DEN-16280_Stuck IN Post Treatment When Inlet Water Alarm Appears
DG-DEN-16280_Stuck IN Post Treatment When Inlet Water Alarm Appears