HD-DEN-7802

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed dialysate temp check for treatment end sub-mode.

Removed dialysate temp check for treatment end sub-mode.

I think we do want to check dialysate temperature during treatment, but maybe not in every sub-mode. For example, I think in treatment end sub-mode we have the dialyzer bypassed (no longer performi...

I think we do want to check dialysate temperature during treatment, but maybe not in every sub-mode. For example, I think in treatment end sub-mode we have the dialyzer bypassed (no longer performing dialysis) so there is no need to worry about the dialysate temp anymore. That is the only exception I can think of at the moment.

HD-DEN-7802_DG-DEV Dialysate Generation
HD-DEN-7802_DG-DEV Dialysate Generation