LEAHI-TD-FIRMWARE-LDT-5624

I don't have a good option. If we're going to change it to pull approach later, can't we just ignore rate changes when doing fluid bolus with sub fluid until you implement the pull approach? Are we...

I don't have a good option. If we're going to change it to pull approach later, can't we just ignore rate changes when doing fluid bolus with sub fluid until you implement the pull approach? Are we even able to do sub fluid bolus yet? Do we have HDF tube sets?

Sure

Sure

I'll implement the pull approach later, this is temporary fix. We need to run Sub pump at Qb rate, if Qb is updated, how can we increase Qsub?

I'll implement the pull approach later, this is temporary fix.
We need to run Sub pump at Qb rate, if Qb is updated, how can we increase Qsub?

Should we check to see if blood sitting timer has not exceeded time limit before we resume (similar to what we do from rinseback stop state before we start an additional)?

Should we check to see if blood sitting timer has not exceeded time limit before we resume (similar to what we do from rinseback stop state before we start an additional)?

So we will command BP every 50ms for saline (okay, but not the pull approach we talked about) and send DD Qsub commands every 50ms for sub fluid (probably not a good idea)?

So we will command BP every 50ms for saline (okay, but not the pull approach we talked about) and send DD Qsub commands every 50ms for sub fluid (probably not a good idea)?

LEAHI-TD-FIRMWARE-LDT-5624_(TD, SW) UI screen goes to main loading screen when pausing fluid bolus in rinseback
LEAHI-TD-FIRMWARE-LDT-5624_(TD, SW) UI screen goes to main loading screen when pausing fluid bolus in rinseback