DEN-17093: Blood Leak Zeroing in Treatment

- Added a new field to message 0x2F00 to log dialysate pump inlet to blood leak detector flush volume

- Removed the blood leak zeroing state enum read

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 26 more files in changeset.
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

DEN-16204: UI: FW: Pure Water mode turning on intermittently.

- Added a new payload to send the current firmware value with the response to a set from UI

- Changed the RO only mode code to not flip the UI RO only status if the firmware rejected the set value and just get the status in the firmware

DEN-16862: Add a command to reset HD during UI software install

- Added a command to request an HD reset during the UI software and while HD is in service mode.

DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Changed the order of the UI response message from firmware so the accept is always the first field.

DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Addressed the code review comments.

    • -8
    • +24
    ./MTreatmentPressureOcclusionData.cpp
    • -9
    • +21
    ./MTreatmentPressureOcclusionData.h
  1. … 25 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 26 more files in changeset.
DEN-16204: UI: FW: Pure Water mode turning on intermittently

- Addressed the code review comments.

  1. … 3 more files in changeset.
DEN-16864: UI Bloodflow message is incomplete

- Updated the payload of the following messages: 0x002C, 0x0005, 0x0008, 0x006C.

  1. … 13 more files in changeset.
DEN-16851: New Alarm for Increased Occlusion Sensitivity - Added the following fields to mssage 0x0009: arterial long filter, venous long filter, blood pump occl long filter, partial occl base line

    • -8
    • +24
    ./MTreatmentPressureOcclusionData.cpp
    • -9
    • +21
    ./MTreatmentPressureOcclusionData.h
  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 598 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 598 more files in changeset.
DENBUG-204 updated the message payload of message 0x0B00 by adding an integer at the end of the structure

DENBUG-204 updated the diagnostics screens qml with the latest dialysate outlet pump broadcast changes

DENBUG-204 aligned the commented out code in the diagnostics qml

  1. … 6 more files in changeset.
DENBUG-204 updated the message payload of message 0x0B00 by adding an integer at the end of the structure

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 55 more files in changeset.
DEN-16256: Updated - UI.P1B : Display UF Rate

Add missing cpp changes for flowData

DEN-16256: Added - UI.P1B : Display UF Rate

Add the variable to the 0x0B00 message model to include UF Current set rate

Setting of the value received from HD for UF Current Set Rate

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 53 more files in changeset.
DEN-15359: Usability: Arterial and Venous Pressure Alarm Limits

- DEN-15387: UI: Updates for treatment parameter entry (A/V alarm window)

- DEN-15388: UI: Updates for in-treatment adjustment of alarm parameter window (pressure alarm screen)

- Monitoring components for the TxHome, Rinseback, BloodPrime

    • -9
    • +29
    ./MTreatmentPressureOcclusionData.cpp
    • -13
    • +28
    ./MTreatmentPressureOcclusionData.h
  1. … 34 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 52 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

- Added the Dialysis subState to TreatmentStates, message 0x0F

  1. … 6 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

- MSG_ID 0x0B

- Added the keepAlive/UI-CheckIn to the list of AckRequired so if HD won't ack UI, the Alarm, "HD Communication Failure" will be displayed.

  1. … 5 more files in changeset.
DEN-14853 UI Out Of Sync With Firmware On Handled Messages

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-13152 : Flow Rate Settings Display "--" or "0" When Treatment is Paused

- Messages 0x0500 and 0x0800 has been updated and the Prescribed Flow Rate is being used.

- The diagnostics screen updated.

- The treatment main screen updated.

- The treatment flow adjustment screen updated.

- the Flows on the commit has 10f39b63ce8a992280ab123f05a5eb44a27c121b has been separated which has to be reverted back since these two messages almost all the time have the same payloads.

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 594 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 33 more files in changeset.
DEN-14147: Add Required Heparin Volume for Treatment to Syringe Installation Screen

- The replacement finder is implemented

- The instruction updater is changed to meed the needs of the replacement finder.

- Next needs to update the instructions regarding the received message.

  1. … 9 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 94 more files in changeset.
DEN-13645: IMP: Updated the Blood Prime message [0x5900]

- It has 3 parameters (UI expected) but recently changed on the FW and sending only 2 which then UI rejects it. now updated to have only 3.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 153 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 74 more files in changeset.
DEN-12262: INT: Code reviews wrap up & support

- Addressed all the code review comments in another branch in here to facilitate the code review process and make the code ready while the actual code owner were busy supporting system for demo.

- Added another condition to the Alarm top/right countdown to hide the timer when FW is not even in the Treatment mode by the HDOpMode.

  1. … 26 more files in changeset.