HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
Setting first iterator to 0 seems safer (in case somebody adds a new enum before what used to be the first enum). Could also create a redundant 0 enum like below and use the "first" enum to initial...

Setting first iterator to 0 seems safer (in case somebody adds a new enum before what used to be the first enum).
Could also create a redundant 0 enum like below and use the "first" enum to initialize a loop iterator safely:
enum list
{
FIRST_TEMP_SENSOR = 0,
TPI_SENSOR = FIRST_TEMP_SENSOR,
TPO_SENSOR,
TD1_SENSOR,
....
}
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Similar code is in other monitoring functions. monitorTemperatureSensors, monitorPressureSensors iterate over their list of sensors using U08. monitorThermistors iterates over thermistors using typ...

Similar code is in other monitoring functions. monitorTemperatureSensors, monitorPressureSensors iterate over their list of sensors using U08. monitorThermistors iterates over thermistors using typedef THERMISTORS_TEMP_SENSORS_T. Changed load cells to use U08 to be consistent with temperature and pressure monitoring.

Should probably change temperature, pressure and load cells to use correct typedef.

Also, it looks like why the typedefs weren't used was in order to set the initial value of the iterator to 0 (i.e, the first sensor in the list). Seems like there should be a value that is the first item in the list, i.e. TEMPSENSORS_FIRST_SENSOR.

Removed.

Removed.

Remove blank line.

Remove blank line.

Why change this to U08?

Why change this to U08?

Add to header and alignment.

Add to header and alignment.

Branch removed. This "bugfix" branch was created in error.

Branch removed. This "bugfix" branch was created in error.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Function header updated. Assignment statements do not require alignment.

Function header updated. Assignment statements do not require alignment.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Assignment alignment.

Assignment alignment.

Assignment alignment in module.

Assignment alignment in module.

Assignment statement alignment

Assignment statement alignment

Assignment statements.

Assignment statements.

Alignment of what?

Alignment of what?

Not clear what you mean.

Not clear what you mean.

Alignment throughout module and function header.

Alignment throughout module and function header.

Alignment

Alignment

Alignment

Alignment

Use get function for override.

Use get function for override.

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off
HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off
DG-DEN-13946_Certain Temperature Sensors Trigger Wrong Alarm
DG-DEN-13946_Certain Temperature Sensors Trigger Wrong Alarm