HD-DEN-14546_Add Option TO Resume Treatment Without Heparin For Empty Syringe Alarm
HD-DEN-14546_Add Option TO Resume Treatment Without Heparin For Empty Syringe Alarm
I resolved the code as you wrote above, per our conversation.

I resolved the code as you wrote above, per our conversation.

HD-DEN-15105_Dialyzer Flip
HD-DEN-15105_Dialyzer Flip
Blaine very recently changed alarm table (and so we changed code) to make it an HD fault. I guess SRS was not updated. Blaine has agreed to change these alarms as I've noted above. He needs to add ...

Blaine very recently changed alarm table (and so we changed code) to make it an HD fault. I guess SRS was not updated.
Blaine has agreed to change these alarms as I've noted above. He needs to add PRS requirements and then update SRS to match.

I am confused. Copying the properties from the SRSHD document, 1/26/2023 SRSHD 919 lists: Alarm: ALARM_ID_HD_VENOUS_BUBBLE_DETECTED ID: 128 Priority: HIGH HD Fault: No DG Fault: No Stop: Yes Clear...

I am confused. Copying the properties from the SRSHD document, 1/26/2023
SRSHD 919 lists:

Alarm: ALARM_ID_HD_VENOUS_BUBBLE_DETECTED
ID: 128
Priority: HIGH
HD Fault: No
DG Fault: No
Stop: Yes
Clear Immediate: No
Non-Recoverable: No
No Resume / No Rinseback / No End Treatment: No, No, No
No Blood Recirculate: Yes
No Dialysate Recirculate: No
Clear Top Only: No

-----------------------------
This table makes it appear that No Resume / No Rinseback / No End Treatment and clear immediate should all be FALSE. Is there an error in the SRS tables?

Agree that fault property should be FALSE. Clear immediate should remain TRUE. No clear should be FALSE. No resume and no rinseback should be TRUE (we do not think user can recover from this alarm)...

Agree that fault property should be FALSE. Clear immediate should remain TRUE. No clear should be FALSE. No resume and no rinseback should be TRUE (we do not think user can recover from this alarm). But no end treatment property should be FALSE (we do want user to be able to go through post-tx process). No blood recirc should be TRUE (we do not want bubble to go to patient), but no dialysate recirc can be FALSE.

HD-DEN-14976_Unable TO Exit Alarm After HD Unit Plugged Back IN Pre Treatment And Blood Priming
HD-DEN-14976_Unable TO Exit Alarm After HD Unit Plugged Back IN Pre Treatment And Blood Priming
HD-DEN-15095_FW System Faults During Treatment Then Returns TO The Standby Screen Reopen
HD-DEN-15095_FW System Faults During Treatment Then Returns TO The Standby Screen Reopen
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Removed 2nd comment.

Removed 2nd comment.

This section will be automatically created.

This section will be automatically created.

A single comment marker (c++ style), aligned with the other comments in the block would be a little more readable.

A single comment marker (c++ style), aligned with the other comments in the block would be a little more readable.

Alignment

Alignment

Comment alignment.

Comment alignment.

UI-DEN-14947_UI [Misc Minor Bugs] - Ready
UI-DEN-14947_UI [Misc Minor Bugs] - Ready
DG-DEN-14999_Alarm Table Updates Part 2
DG-DEN-14999_Alarm Table Updates Part 2