DEN-14497: Changed CPo and RR alarms to only occur at end of fill based on average conductivity.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14548 changed the time for heat disinfect cooling from 3 hours to 5 hours

Why change this to U08?

Why change this to U08?

DG-DEN-14548_Alarm 247 ON DG Heat Disinfect
DG-DEN-14548_Alarm 247 ON DG Heat Disinfect
DEN-14548 increased the heat disinfect cooling timeout from 90 minutes to 180 minutes

Add to header and alignment.

Add to header and alignment.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Use get function for override.

Use get function for override.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/dg/dgfirmware into develop

Merge branch 'DEN-14086-alarm-159-dialysate-inlet-pump-failed-flow-vs-motor-speed-check' into develop

DEN-14086: fixed logging issue with LC drift alarm.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Fixed merge conflicts.

    • -4
    • +12
    /firmware/App/Controllers/LoadCell.c
DEN-13946 - Correct monitor functions where multiple sensors are "or'd" into a single alarm.

    • -9
    • +13
    /firmware/App/Controllers/LoadCell.c
    • -8
    • +12
    /firmware/App/Controllers/Pressures.c
    • -4
    • +16
    /firmware/App/Controllers/Thermistors.c
DEN-14086: Increase loadcell drift allowance from +/- 60 to 80 grams - and for heat disinfect, +/- 200 grams.

    • -3
    • +11
    /firmware/App/Controllers/LoadCell.c
DG-DEN-13946_Certain Temperature Sensors Trigger Wrong Alarm
DG-DEN-13946_Certain Temperature Sensors Trigger Wrong Alarm
DG-DEN-14086_Alarm 159 Dialysate Inlet Pump Failed Flow VS Motor Speed Check
DG-DEN-14086_Alarm 159 Dialysate Inlet Pump Failed Flow VS Motor Speed Check
Assume this is for testing?

Assume this is for testing?

DEN-14497 updated the drain pump control for the flow and updating chemical disinfect is in progress

    • -10
    • +19
    /firmware/App/Controllers/DrainPump.c
    • -161
    • +113
    /firmware/App/Modes/ModeChemicalDisinfect.c
    • -5
    • +3
    /firmware/App/Modes/ModeHeatDisinfect.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/Thermistors.c
    • -1
    • +1
    /firmware/App/Controllers/UVReactors.c