DEN-15246 addressed the code review comments

Done

Done

Is this a big risk? At this time I am not too concerned about it. If you think it should be addressed, I can make changes.

Is this a big risk? At this time I am not too concerned about it. If you think it should be addressed, I can make changes.

Done

Done

Done

Done

This is 105.

This is 105.

DEN-15274 work in progress, added a set of commands to be able to get, send, and set the load cells tare values

    • -0
    • +40
    /firmware/App/Controllers/LoadCell.c
DG-DEN-15274_HD DG Test Configuration Setup
DG-DEN-15274_HD DG Test Configuration Setup
DG Staging Unreviewed Commits
DG Staging Unreviewed Commits
DG-DEN-14859_Cleaning Modes State Timer Reset
DG-DEN-14859_Cleaning Modes State Timer Reset
Merge branch 'staging' of ssh://dvm-linux02:7999/dg/dgfirmware into staging

Updated HAL

    • -138
    • +94
    /firmware/source/sys_startup.c
DEN-16565 cherry picked Sunrise commit e3de58ef4bdffc57c40f661dded0040055971e1e by Dara Navaei

    • -138
    • +94
    /firmware/source/sys_startup.c
Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Modes/ModeHeatDisinfect.c
Merged DEN-15265

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-15265 addressed a code review comment

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

name is idleState (not state).

name is idleState (not state).

DEN-15265 addressed a code review comment

Should be &result.

Should be &result.

DEN-15265, DEN-15236 changed the chemical disinfect monitor to check the dialysate cap not to be closed in during the disinfect since the PAA acid is connected to it.

Bamboo Commit: Updated DGCommon.h with build versions from Bamboo

Merged DEN-15265

DEN-15265, DEN-15246 changed the heat disinfect's primary target temperature to 60 C prior to filling with water for the actual heat disinfct. Updated the code to properly exit the bad fill state of the gen idle mode once the bad sequence is finished

    • -1
    • +2
    /firmware/App/Modes/ModeHeatDisinfect.c