Controllers

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-9906: Fixed doxygen comments and group for dialysate flow module.

Merged staging and resolved merge conflicts

  1. … 21 more files in changeset.
DEN-9906 Added dialysate flow rate out of range alarm

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Merge branch 'staging' of ssh://dvm-linux02:7999/dg/dgfirmware into staging

  1. … 1 more file in changeset.
Removed the DEBUG_DENALI comment

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 78 more files in changeset.
DEN-9480 updated documentation

  1. … 1 more file in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

The const is an optional step in our code. Plus these are the local variables and their values are not changed inside the function.

The const is an optional step in our code. Plus these are the local variables and their values are not changed inside the function.

Done.

Done.

Done.

Done.

Variable reservoirDataPublicationTimerCounter is missing in the @detail Inputs and Output

Variable reservoirDataPublicationTimerCounter is missing in the @detail Inputs and Output

I am not sure why you removed the "const" from dialysateTemp and targetTemp?

I am not sure why you removed the "const" from dialysateTemp and targetTemp?

DEN-9906 added self test for the dialysate flow sensor and removed its broadcast function

  1. … 3 more files in changeset.
Added "pressuresDataPublicationTimerCounter" variable to @details Outputs

Added "pressuresDataPublicationTimerCounter" variable to @details Outputs

The state variable valvesStatesPublicationTimerCounter was read, increment, then write. Therefore, it needs to be added to the @details Outputs.

The state variable valvesStatesPublicationTimerCounter was read, increment, then write. Therefore, it needs to be added to the @details Outputs.

Merged dialysate flow sensor calibration

  1. … 2 more files in changeset.
DEN-9906 Removed code added to collect unfiltered dialysate flow data

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Need to range check flow.

Need to range check flow.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

I removed the alarm.

I removed the alarm.