DEN-11928 Updated conductivity

DEN-11928 Updated conductivity test

  1. … 1 more file in changeset.
DEN-11928 Updated temperature out of range test

  1. … 1 more file in changeset.
DEN-11928 Updated temperature, rejection ratio, CPo with new values

  1. … 1 more file in changeset.
DEN-12225 updated the UV reactors

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' into develop

  1. … 1 more file in changeset.
DEN-11750 updated the code

  1. … 2 more files in changeset.
DEN-11928 Implemented conductivity updates

  1. … 1 more file in changeset.
Could we move these fields outside of the for loop? They dont need to be initialized NUM_OF_FANS_NAMES times.

Could we move these fields outside of the for loop? They dont need to be initialized NUM_OF_FANS_NAMES times.

If the DG is broadcasting these raw values, shouldn't Dialin be sync'd as well to handle these new fields?

If the DG is broadcasting these raw values, shouldn't Dialin be sync'd as well to handle these new fields?

All the new timer counters no longer are initialized to 0. I understand that they get set in their respective init()'s to the timer counter, but was this intentional?

All the new timer counters no longer are initialized to 0. I understand that they get set in their respective init()'s to the timer counter, but was this intentional?

We should probably keep the = 0;

We should probably keep the = 0;

Merge branch 'DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up' of ssh://dvm-linux02:7999/dg/dgfirmware into DEN-11750-dg_dev-dg-dialysate-temperature-control-tune-up

DEN-11750: Fixed casting error in ConcentratePumps.c.

Merged DEN-11750 and resolved the coflicts

  1. … 3 more files in changeset.
Most of the files are VectorCAST that has been brought back from Code Clinic. Look at the files but they are auto generated so they are mostly passing through.

Most of the files are VectorCAST that has been brought back from Code Clinic. Look at the files but they are auto generated so they are mostly passing through.

Notices the 170 files that need to be reviewed https://c.tenor.com/RXMxqsRKEn0AAAAC/this-is-fine.gif

Notices the 170 files that need to be reviewed

https://c.tenor.com/RXMxqsRKEn0AAAAC/this-is-fine.gif

DEN-11750 addressed the bugs that Code Clinic reported

  1. … 3 more files in changeset.
DG-DEN-11750_DG Dev Dialysate Temperature Control Tune Up
DG-DEN-11750_DG Dev Dialysate Temperature Control Tune Up
DEN-11750 added more comments

DEN-11750 worked on the RO pump duty cycle feedback

  1. … 2 more files in changeset.
DEM-11750 updated the code

  1. … 1 more file in changeset.
DEN-11750: Fixed drain pump RPM out of range alarm.

DEN-11750 updated the drain pump

  1. … 1 more file in changeset.
Updated code to store the raw conductivity data from sensors CD1 and CD2

DEN-11750 updated the drain pump RPM to make sure the pump's RPM is only checked if the pumpis running

DEN-11250 Updated conductivity sensors manufacture information

DEN-11750 Fixed temperature compensated conductivity error and added code to publish raw conductivity data. Updated FW code per peer review comments

DEN-11750 enabled the load cells tare alarm

DEN-11750 Fixed temperature compensated conductivity error and added code to publish raw conductivity data