set the RO pump rate to setROPumpTargetFlowRateLPM( getTargetFillFlowRateLPM(), TARGET_RO_PRESSURE_PSI );
Updated code to check for the difference in percent of conductivity between CD1 and CD2 and make sure it is < 5% otherwise alarm.
Uncommented and recompiled without errors
RESOLVED in CODE WALKTHROUGH.
Alarm is recoverable. Should go to fill paused state. Paused state should go back to this test inlet water quality state when alarm acknowledged.
Added ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY to AlarmDefs.h Uncommented initPersistentAlarm(ALARM_ID_INLET_WATER_HIGH_CONDUCTIVITY
Let's review requirement - may be obsolete.
Let's review that requirement - may be obsolete.