DEN-14307: Added support for parking concentrate pumps.

  1. … 12 more files in changeset.
DEN-14307 updated the cleaning modes per the SRS changes

  1. … 7 more files in changeset.
Merge branch 'DEN-14433-sw-s86-mg-publish-fixes' into develop

DEN-14433: DIAL-123: fixed broadcast messages to send if we are >= of their respective intervals instead of >.

Merge branch 'DEN-13050-vpi-does-not-close-after-sample-water-returns-to-flush-idle' into develop

DEN-13757 - Remove unused function monitorLoadCellsWeightOutOfRange/

DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DEN-13757 - Correction to load cell alarm.

DEN-13757 - All 4 load cells have a single persistent alarm. Change to correspond to how pressure alarms are handled.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14253 addressed the code review comments

Done.

Done.

Done.

Done.

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

Merged DEN-14253

  1. … 7 more files in changeset.
DEN-14253 updated the code so the trimmer heater with no flow uses the raw dialysate flow prior to applying the calibration to it.

  1. … 1 more file in changeset.
DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
DEN-14403-dg-voltages-alarm-77-not-triggered FPGA 1V alarm limits modified to +/-5% from +/-10%, to match requirements