DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Done.

Done.

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

I renamed the #define to be consistent with the latest changes.

I renamed the #define to be consistent with the latest changes.

Seems like this is a minimum (not a maximum). Pressure below this level is an alarm which enforces a minimum pressure of 14 PSI.

Seems like this is a minimum (not a maximum). Pressure below this level is an alarm which enforces a minimum pressure of 14 PSI.

DG-DEN-14356_Change DG Inlet Pressure Limit
DG-DEN-14356_Change DG Inlet Pressure Limit
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Correct. Fixed. Thanks!

Correct. Fixed. Thanks!

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.