DEN-14307 added a command to be able to start and stop heat disinfect active cool mode. Added a monitor function to check the status of the caps during the heat disinfect active cool mode. Updated heating code to use TD2 temperature sensor to set the primary heater's target for the later DVT devices

  1. … 10 more files in changeset.
DEN-14307: Added support for parking concentrate pumps.

  1. … 12 more files in changeset.
DEN-14307 updated the cleaning modes per the SRS changes

  1. … 7 more files in changeset.
Merge branch 'DEN-14433-sw-s86-mg-publish-fixes' into develop

DEN-14433: DIAL-123: fixed broadcast messages to send if we are >= of their respective intervals instead of >.

Merge branch 'DEN-13050-vpi-does-not-close-after-sample-water-returns-to-flush-idle' into develop

DEN-13757 - Remove unused function monitorLoadCellsWeightOutOfRange/

DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DG-DEN-13050_Vpi Does Not Close After Sample Water Returns TO Flush Idle
DEN-13757 - Correction to load cell alarm.

DEN-13757 - All 4 load cells have a single persistent alarm. Change to correspond to how pressure alarms are handled.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14253 addressed the code review comments

Done.

Done.

Done.

Done.

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

Can't we just say isFlowLow = ( measFlow > minFlow ? FALSE : TRUE );

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

For both sensors, the .data should only be set once to final value. Use a temporary local variable until it's ready.

Merged DEN-14253

  1. … 7 more files in changeset.
DEN-14253 updated the code so the trimmer heater with no flow uses the raw dialysate flow prior to applying the calibration to it.

  1. … 1 more file in changeset.
DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14253_Dialysate Flow Too Low Alarm Occurs AT 100 ML Min
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
DG-DEN-14403_DG Voltages Alarm 77 Not Triggered
DEN-14403-dg-voltages-alarm-77-not-triggered FPGA 1V alarm limits modified to +/-5% from +/-10%, to match requirements