DEN-14646 Add window timer support for communications alarms. Debug code for testing conductivity sensor FPGA generated alarms.

  1. … 1 more file in changeset.
DEN-14788: Fixed heaters duty cycle override to immediately set PWM instead of waiting for control state machine to set PWM.

Add function header.

Add function header.

Add function headers.

Add function headers.

Add tempBelowTargetStartTimeMS and rsrvrsStatus.

Add tempBelowTargetStartTimeMS and rsrvrsStatus.

Add function headers.

Add function headers.

Add comments and alignment.

Add comments and alignment.

The revised comment doesn't make sense. What creates the flow exactly?

The revised comment doesn't make sense. What creates the flow exactly?

DG-DEN-14788_SW_S91_Dial-222_DEN-14831
DG-DEN-14788_SW_S91_Dial-222_DEN-14831
DEN-14763, DIAL-233, DEN-14812 added a command to be able to set the dialysate mixing ratios and added an event to send the average conductivity of acid and bicarb at the end of the fill. Changed the measured flow function to flow with concentrate pumps in drain pump's control to flow function

  1. … 7 more files in changeset.
Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DG-DEN-14788_SW S91 MG CP Persistence
DG-DEN-14788_SW S91 MG CP Persistence
Merge branch 'DEN-14789-sw-cp-persistence-timer-increase' into develop

  1. … 3 more files in changeset.
DEN-14789: Increased CP measured speed persistence to 6s in order to handle pump parking

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14781-alarm-117-dg-can-message-not-acked' into staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-14649: Fixed park fault to consider override

DEN-14649: fixed conc pump park faults to no longer use measured speed.

DEN-14781 changed the drain pump flow control interval and the trimmer heaters minimum flow rate

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

There is not a #define for 0.0F.

There is not a #define for 0.0F.

Should be #define

Should be #define

DEN-14781 updated the code to be able to support Nelson modes

  1. … 4 more files in changeset.
DG-DEN-14689_Messaging Bugs_DGFW_and_Common
DG-DEN-14689_Messaging Bugs_DGFW_and_Common