DEN-14646 Add window timer support for communications alarms. Debug code for testing conductivity sensor FPGA generated alarms.

  1. … 1 more file in changeset.
DEN-14788: Fixed heaters duty cycle override to immediately set PWM instead of waiting for control state machine to set PWM.

DEN-14763, DIAL-233, DEN-14812 added a command to be able to set the dialysate mixing ratios and added an event to send the average conductivity of acid and bicarb at the end of the fill. Changed the measured flow function to flow with concentrate pumps in drain pump's control to flow function

  1. … 7 more files in changeset.
Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14789-sw-cp-persistence-timer-increase' into develop

  1. … 3 more files in changeset.
DEN-14789: Increased CP measured speed persistence to 6s in order to handle pump parking

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14781-alarm-117-dg-can-message-not-acked' into staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-14649: Fixed park fault to consider override

DEN-14649: fixed conc pump park faults to no longer use measured speed.

DEN-14781 changed the drain pump flow control interval and the trimmer heaters minimum flow rate

DEN-14781 updated the code to be able to support Nelson modes

  1. … 4 more files in changeset.
DEN-14781 updated Nelson support

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-14646-fpga-alarm-properties-and-timing-3

DEN-1646 - Update for conductivity sensor alarms.

DEN-14646 - Remove checking the error counter for persistent alarms that previously generated alarms based upon the read counter and the error counter.

Merge branch 'DEN-14589-DIAL-211' into develop

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14760 changed the minimum dialysate flow rate of the trimmer heater to alarm.

DEN-14589: Implemented concentrate pump fault persistence timer.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-13972 added a #define for the delta temperature in the heaters

DEN-13972 changed the code in the heaters to consider if the temperatures are negative and therefore the duty cycle is not set to positive artificially

Fixed merge conflict.

  1. … 4 more files in changeset.
Fixed merge conflicts.

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
DEN-14604 added VectorCAST from PiSquare and addressed code review comments

  1. … 2 more files in changeset.