DEN-15973 Change #define name

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting...

Consider calling a (new) function in ConcentratePumps.c that would do all of this (decides whether to accept or reject command and parks the pumps if accepting, sets reject reason code if rejecting). Function would also create and send the cmd response message and return TRUE/FALSE which would set local result here for the ack response message below.
We should reject cmd if pumps are busy/on and/or if current DG mode is fill or fault mode.

DEN-15973 Correct #define name

DEN-13973 Rename alarm

DEN-15973 Add RO Pump max PWM alarm

DG-DEN-15973_Alarm 48 UF Rate Tare Error Triggered During Treatment
DG-DEN-15973_Alarm 48 UF Rate Tare Error Triggered During Treatment
DG-DEN-16093_Alarm 97 Conductivity Out OF Range Triggered ON First Fill OF Treatment
DG-DEN-16093_Alarm 97 Conductivity Out OF Range Triggered ON First Fill OF Treatment
Merged DEN-16250

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DG-DEN-15974_Alarm 187 Triggered During Consumables Check And During Treatment
DG-DEN-15974_Alarm 187 Triggered During Consumables Check And During Treatment
Merged DEN-15964

  1. … 4 more files in changeset.
DEN-15974 removed the clearance time of the persistence alarms of the concentrate pumps

Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DG-DEN-16212_DG Flush Happy Path Failed
DG-DEN-16212_DG Flush Happy Path Failed
DEN-16212 fixed the concentrate pumps speed alarm

Merged DEN-16149

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-16125 updated the concentrate pumps code to set the direction speed of the concentrate pumps when the measured speed is called

  1. … 2 more files in changeset.
DG-DEN-16125_DG Chem Flush DG Does Not Transition From Flush R1 TO R2 Drain R2 TO Cancel Water
DG-DEN-16125_DG Chem Flush DG Does Not Transition From Flush R1 TO R2 Drain R2 TO Cancel Water
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.