DEN-15973 Change to alarm limits and persistence values

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Why 0.0F? memset() doesn't care about type - it just sets a number of bytes to given value.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
Please remove the blank line.

Please remove the blank line.

DEN-15973 Move max pwm check to execROPumpMonitor

Merged DEN-16309

  1. … 2 more files in changeset.
DG-DEN-15791_DG Software Design Description
DG-DEN-15791_DG Software Design Description
Merge branch 'DEN-15973-alarm-48-uf-rate-tare-error-triggered-during-treatment' into develop

# Conflicts:

# firmware/App/DGCommon.h

  1. … 2 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-16309 added a new heat disinfect state

  1. … 2 more files in changeset.
DEN-16309 removed the clearance period some of the persistence alarms

The ROPump PWM is set from 5 different calls. Doesn't look like there is a global indicating the last set PWM value (could add). The execROPumpMonitor function might be a good place to put the chec...

The ROPump PWM is set from 5 different calls. Doesn't look like there is a global indicating the last set PWM value (could add). The execROPumpMonitor function might be a good place to put the check as it looks to be able to handle different modes, but again doesn't have access to the last set PWM value.

Merge branch 'staging' into DEN-15973-alarm-48-uf-rate-tare-error-triggered-during-treatment

DEN-15973 Address code review comments

Corrected

Corrected