DEN-14175 addressed the code review comments and replaced the V3 enable software configuration flag with the hardware configuration flag from FGPA. Updated the trimmer heater's ramp state to check the use previous duty cycle flag instead of estimating the duty cycle.

  1. … 15 more files in changeset.
This is done to get an average of the fill temperature as well as the last fill temperature prior to tranistioning to Gen Idle. These values are used for the next estimation and are different from ...

This is done to get an average of the fill temperature as well as the last fill temperature prior to tranistioning to Gen Idle. These values are used for the next estimation and are different from the heaters driver.

Done.

Done.

No, it should be removed.

No, it should be removed.

DEN-14237 Correct getHeatDisinfectStatus logic.

DEN-14237 DG_HEAT_DISINFECT_STATE_MIX_DRAIN_R2 now transitions to DG_HEAT_DISINFECT_STATE_COMPLETE.

Where is valveSensedStates being updated?

Where is valveSensedStates being updated?

Remove comment.

Remove comment.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

Not clear this needs to be done. I believe is properly handled in heaters when you get a temperature value and configured for V3.

DEN-14232 updated the VPo status to prevent filling the reservoirs at the end of heat disinfect

DEN-14175, DIAL-213 added drain pump current and direction override commands, updated the heaters driver with the latest optimization requirements, added a DG valves sensed states override command

  1. … 8 more files in changeset.
DEN-14232 Turn off trimmer heater before switching to FILL_R2_WITH_HOT_WATER and then back on when switching to HEAT_DISINFECT_R2_TO_R1.

DG-DEN-14232_Heat Disinfect Alarm 199 During Fill R2 With Hot Water
DG-DEN-14232_Heat Disinfect Alarm 199 During Fill R2 With Hot Water
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
DG-DEN-14175_DG HD Dev HD DG Dvt Update Part 5
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-14101: Fixed warnings on implicit declaration.

Merge branch 'staging' into DEN-13786-hd_dg_dev-s78_bb

  1. … 1 more file in changeset.
Merge branch 'DEN-14101-sw-s82-mg-DIAL-179-DIAL-184' into develop

  1. … 5 more files in changeset.
DEN-14101: DIAL-184: Fixed publish interval to not overlap with the publish counter.

DEN-14101: DIAL-184: Fixed gen idle/bad fill publish override handler to actually work.

  1. … 3 more files in changeset.
DEN-14101: DIAL-184: reworked bad fill broadcast to now be genidle broadcast containing bad fill states and signal.

DG-DEN-14166_DG HD Dev HD DG Dvt Update Part 4
DG-DEN-14166_DG HD Dev HD DG Dvt Update Part 4
DEN-13786 Correct VPo