Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15571, DEN-15398 changed the fill mode full speed prime time from 8 seconds to 7 seconds

DEN-15571 addressed the code review comments

Done

Done

DEN-15571 updated the chemical disinfect drain status function

Several variables missing from header.

Several variables missing from header.

DEN-15571 made changes in the chemical disinfect mode to not trigger the max temeprature and conductivity out of range until the end of the chemical disinfect mode

DG-DEN-15571_Chemical Disinfect RX TO RY Timeout Alarms DO Not Trigger
DG-DEN-15571_Chemical Disinfect RX TO RY Timeout Alarms DO Not Trigger
Merged staging

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

DEN-15571 changed the chemical disinfect temperature timeout from 1 minute to 5 minutes

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 4 more files in changeset.
DEN-15367 removed the payload from the message that informs HD that DG is ready to dispense more fluid in DG. DEN-15571 fixed the chemical disinfect state machine to transition to cancel water path once the target temperature was not reached within the specified time

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Still has a payload. Should remove it.

Still has a payload. Should remove it.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed the check to make sure the target speed is > 0. The current speed is always >= 0.0 mL/min.

Removed the check to make sure the target speed is > 0. The current speed is always >= 0.0 mL/min.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH