Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15614 matched the cleaning modes with the SRS(s) DEN-15978 added more comments to the switches driver

  1. … 1 more file in changeset.
DEN-15614 changed the cleaning modes to request a transition to fault mode if a fault alarm is triggered

  1. … 1 more file in changeset.
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DG-DEN-15614_New Updated Disinfect Actuators Requirement Issues List
DEN-15614 updated the cleaning modes actuators with the SRSs

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-15921 code clean up

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15921, DEN-15916 changed the heat disinfect, heat disinfect active cool, and flush modes to send their start state via the events system

  1. … 1 more file in changeset.
DEN-15921 addressed a code review comment

Done thanks.

Done thanks.

DEN-15921, DEN-15882 made a change in the code to transition to drain R1 in RO permeate sample mode after the sample collection is finished

This comment should say "bicarb pump is 2% faster than...." .

This comment should say "bicarb pump is 2% faster than...." .

Merge branch 'DEN-15921-when-cap-is-open-dg-transition-to-cancel-basic-path-instead-of-cancel-water-path' into develop

  1. … 1 more file in changeset.
DEN-15921, DEN-15868 changed the direction of the pumps in RO permeate sample mode. Changed the code to check the flow in heat disinfect and active cool to only when the system is not hot

  1. … 3 more files in changeset.
DEN-15517 added an event to monitor the inactive data

DEN-15921 changed the active cool to transition to cancel water path instead of basic path when caps are not on

DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
DG-DEN-15921_When Cap IS Open DG Transition TO Cancel Basic Path Instead OF Cancel Water Path
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
DEN-15736 updated the concentrate pumps to set the control bit if it is not the same as what it should be

  1. … 3 more files in changeset.
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test
DG-DEN-15736_Concentrate Pump Runs IN Opposite Direction During Concentrate Self Test
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DG-DEN-15867_System Does Concentrate Self Test Twice During Pre Treatment
DG-DEN-15867_System Does Concentrate Self Test Twice During Pre Treatment
DEN-15867: Fix for issue where concentrate prime/test was being done twice in pre-tx.